RWTH Aachen University needs an accelerated rate calorimeter supplied to the Städteregion Aachen for battery-cell and comparable electrochemical energy-storage safety testing. The equipment must support testing at temperatures of at least 200°C, accommodate cells of at least 20 Ah or test specimens at least 100 mm long, and provide at least four temperature or equivalent safety-relevant measurement channels with documented data acquisition; dynamic abuse tests require a rate of at least 100 Hz or equivalent. Its safety concept must include at least three specified safeguards, such as temperature and pressure monitoring, pressure relief, controlled gas exhaust, inerting or automatic shutdown. Suppliers must document one verifiable reference project commissioned within the past 10 years for comparable equipment, with a European customer and reference contact details. The required public-liability insurance cover is at least €1 million per year, combined for property damage, personal injury and financial loss.